If the DAC UNIVERSAL shows an error in the display, it is important to find out exactly, what this error means and what
action to take. Below you will find a list of the error codes.
If the DAC UNIVERSAL displays an error, you cannot be sure that a complete cycle has been carried out and therefore the
instruments are not sterile – so you must start a complete new cycle.
If an error occurs during sterilization, the display will, when the lid opens, show that an error has occurred during
sterilization. In order to clear this error and start a new complete cycle — you need to press the «M»-button together with the
«C»-button (This will also be indicated in the display)
If the error light is flashing without giving an alarm signal, it is because the machine has reached 3000 cycles or it has
been 24 months since the last service. Contact your supplier in order to do a biennial service on the DAC UNIVERSAL
autoclave. Meanwhile the machine can still be used if „only‟ the error light is flashing and no other error codes are visual in
the display.
In case of an error, an alarm will sound and the error will be displayed visually.
Please note the error code before resetting the alarm. Pressing C resets the alarm.
When pressing C the lid will for approx. 2 seconds raise and then stop, until the temperature is below 104C/219.2F,
following which it will open completely.
It is, however, possible to force the lid to open completely by pressing C repeatedly. For every press on C, the lid will
continue opening for 2 seconds. We do not recommend this feature to be used.
NB: Do not open if the temperature is more than 104C/219.2F (C indicated in the display), hot steam might come out of
the chamber.
Error
64
65
66
67
68
70
71
72
73
74
75
76
77
78
79
81
82
83
84
34
DAC UNIVERSAL – USER MANUAL
729050 — Version 16 S
Description
The safety system has been activated Press C Restart the autoclave.
The safety system is defective Please contact your supplier.
The temperature during sterilization is too high Please contact your supplier.
The time of sterilization has been less than 3 minutes The instruments should
be sterilised again Please contact your supplier if this error appears more than
once.
The pressure is too high during pressure test Please contact your supplier
The pressure level does not decrease fast enough after pressure test Empty
the chamber for water Clean or change the filter Check that the drain tube is
not blocked or broken (note that the drain tube must be of heat resistant material,
e.g. PTFE) Empty the drain tank if full Please contact your supplier.
The autoclave was blocked due to use of bad water Control the level of micro-
siemens in the water Empty the watertank Fill in fresh water (below 3S)
The closing system was blocked on the way up Remove possible blocking for
the lift Please contact your supplier
The sterilization temperature has not been reached fast enough before the first
back-flush Pour in water in case of low level Please contact your supplier
The right temperature for building up pressure for sterilization has not been
reached fast enough Pour in water if low level If the autoclave is placed in a
cold spot remove it to surroundings at room temperature Press C to reset and
then restart Please contact your supplier
The right sterilization temperature has not been reached fast enough Pour in
water if low level If the autoclave is placed in a cold spot move it to a
surrounding at room temperature Press C to reset and then restart Please
contact your supplier
Valve error Please contact your supplier or try with only 5 instruments on the lid.
The autoclave does not lubricate Please contact your supplier
Valve error Please contact your supplier or try with only 5 instruments on the lid.
Valve error Please contact your supplier or try with only 5 instruments on the lid.
Low level in either the water tank or the oil bottle Pour water into the water tank
or change the oil bottle Please contact your supplier
The temperature in the steam generator is too high Please contact your
supplier
Temperature in chamber is too high Please contact your supplier
The sterilization temperature has not been maintained during sterilization Pour
in water, if needed Please contact your supplier
60 53 362 D 3472
D3472.201.01.09.02
Стерилизатор паровой комбинированный DAC UNIVERSAL ф.Sirona |
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
If the DAC UNIVERSAL shows an error in the display, it is important to find out exactly, what this error means and what
action to take. Below you will find a list of the error codes.
If the DAC UNIVERSAL displays an error, you cannot be sure that a complete cycle has been carried out and therefore the
instruments are not sterile – so you must start a complete new cycle.
If an error occurs during sterilization, the display will, when the lid opens, show that an error has occurred during
sterilization. In order to clear this error and start a new complete cycle — you need to press the «M»-button together with the
«C»-button (This will also be indicated in the display)
If the error light is flashing without giving an alarm signal, it is because the machine has reached 3000 cycles or it has
been 24 months since the last service. Contact your supplier in order to do a biennial service on the DAC UNIVERSAL
autoclave. Meanwhile the machine can still be used if „only‟ the error light is flashing and no other error codes are visual in
the display.
In case of an error, an alarm will sound and the error will be displayed visually.
Please note the error code before resetting the alarm. Pressing C resets the alarm.
When pressing C the lid will for approx. 2 seconds raise and then stop, until the temperature is below 104C/219.2F,
following which it will open completely.
It is, however, possible to force the lid to open completely by pressing C repeatedly. For every press on C, the lid will
continue opening for 2 seconds. We do not recommend this feature to be used.
NB: Do not open if the temperature is more than 104C/219.2F (C indicated in the display), hot steam might come out of
the chamber.
Error
64
65
66
67
68
70
71
72
73
74
75
76
77
78
79
81
82
83
84
34
DAC UNIVERSAL – USER MANUAL
729050 — Version 16 S
Description
The safety system has been activated Press C Restart the autoclave.
The safety system is defective Please contact your supplier.
The temperature during sterilization is too high Please contact your supplier.
The time of sterilization has been less than 3 minutes The instruments should
be sterilised again Please contact your supplier if this error appears more than
once.
The pressure is too high during pressure test Please contact your supplier
The pressure level does not decrease fast enough after pressure test Empty
the chamber for water Clean or change the filter Check that the drain tube is
not blocked or broken (note that the drain tube must be of heat resistant material,
e.g. PTFE) Empty the drain tank if full Please contact your supplier.
The autoclave was blocked due to use of bad water Control the level of micro-
siemens in the water Empty the watertank Fill in fresh water (below 3S)
The closing system was blocked on the way up Remove possible blocking for
the lift Please contact your supplier
The sterilization temperature has not been reached fast enough before the first
back-flush Pour in water in case of low level Please contact your supplier
The right temperature for building up pressure for sterilization has not been
reached fast enough Pour in water if low level If the autoclave is placed in a
cold spot remove it to surroundings at room temperature Press C to reset and
then restart Please contact your supplier
The right sterilization temperature has not been reached fast enough Pour in
water if low level If the autoclave is placed in a cold spot move it to a
surrounding at room temperature Press C to reset and then restart Please
contact your supplier
Valve error Please contact your supplier or try with only 5 instruments on the lid.
The autoclave does not lubricate Please contact your supplier
Valve error Please contact your supplier or try with only 5 instruments on the lid.
Valve error Please contact your supplier or try with only 5 instruments on the lid.
Low level in either the water tank or the oil bottle Pour water into the water tank
or change the oil bottle Please contact your supplier
The temperature in the steam generator is too high Please contact your
supplier
Temperature in chamber is too high Please contact your supplier
The sterilization temperature has not been maintained during sterilization Pour
in water, if needed Please contact your supplier
60 53 362 D 3472
D3472.201.01.09.02
Стерилизатор паровой комбинированный DAC UNIVERSAL ф.Sirona |
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
Specifications:846/846056-dac_universal.pdf file (02 Feb 2023) |
Accompanying Data:
Sirona DAC UNIVERSAL Dental equipment, Medical Equipment PDF Operating Instruction (Updated: Thursday 2nd of February 2023 01:13:31 PM)
Rating: 4.8 (rated by 10 users)
Compatible devices: ORTHOPHOS XG 5, C3 Plus, inEos Blue, SIDEXIS XG, c8+ turn, CEREC MC XL Premium Package, DAC PROFESSIONAL, Sinius.
Recommended Documentation:
Text of Operating Instruction
(Ocr-Read Version Summary of Contents, UPD: 02 February 2023)
-
198, 60 53 362 D3472 198 D3472.201.10.20.02 07.2016 9Cleaning and maintenance DAC UNIVERSAL STANDARD Sirona Dental A/S 9.6Cleaning the water tank Operating Instructions DAC UNIVERSAL 9. Press the «Menu/Enter»-Key to confirm the selection. 10. Press the «Up» or «Down»-Key until «Draintank» is selected. 11. Press the «Menu/Enter»-Key to conf…
-
5, 60 53 362 D3472 D3472.201.10.20.02 07.2016 5 Sirona Dental A/S Table of Contents Operating Instructions DAC UNIVERSAL båÖäáëÜ 5.1.2.1 Testing with a chemical indicator……………………………….. 60 5.1.2.2 Testing with a PCD process challenge device …………….. 64 5.1.2.3 Testing with a protein residue test……………………………… 66 …
-
31, 60 53 362 D3472 D3472.201.10.20.02 07.2016 31 Sirona Dental A/S 3Product description Operating Instructions DAC UNIVERSAL 3.2Controls and functional elements of the DAC UNIVERSAL båÖäáëÜ 3.2.4 Back of unit See also «Installation of DAC UNIVERSAL connections [ → 49]». I: «Wrapped»-Key The «Wrapped»-Key is for the cycle for reprocessing …
-
188, 60 53 362 D3472 188 D3472.201.10.20.02 07.2016 8Operation DAC UNIVERSAL STANDARD Sirona Dental A/S 8.3Reprocessing with DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL General attachment of scaling tips 1. The scaler tip was unscrewed from the ultrasonic handpiece with the torque wrench and the tip remains in the torque wrench. 2. Ensure that the torque wre…
-
141, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 141 Sirona Dental A/S 7Display messages DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL båÖäáëÜ IMPORTANT If a display message is displayed during sterilization, it is shown again when the lid opens. To delete this display message and start a new cycle, you must press the «Up»-Key and the «Right»-Key simultan…
-
155, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 155 Sirona Dental A/S 8Operation DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL 8.1Regular activities båÖäáëÜ The appropriate indicator holder is used with a Class 5 chemical indicator, when using the standard or sensitive program. The appropriate indicator holder is used with a Class 6 chemical indicator, …
-
49, 60 53 362 D3472 D3472.201.10.20.02 07.2016 49 Sirona Dental A/S 4Installation Operating Instructions DAC UNIVERSAL 4.1Requirements of the location / Ambient conditions båÖäáëÜ 4 Installation 4.1 Requirements of the location / Ambient conditions Requirements To find a suitable location for the DAC UNIVERSAL, observe the following requirements: ● ● Open, well-ventilated ar…
-
84, Sirona DAC UNIVERSAL 60 53 362 D3472 84 D3472.201.10.20.02 07.2016 5Operation DAC UNIVERSAL ADVANCED Sirona Dental A/S 5.2Using the menu and display information DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL 5.2.2.3.1 Program selection The program selection menu enables the user to select the appropriate program. Please refer to section „Programs and program selection [ → 33]“ for details.…
-
12, 60 53 362 D3472 12 D3472.201.10.20.02 07.2016 1General Information Sirona Dental A/S 1.7Warranty and liability Operating Instructions DAC UNIVERSAL 1.7 Warranty and liability The manufacturer guarantees that the product delivered to you (DAC UNIVERSAL) was subjected to a thorough quality test before it left the factory. A warranty period of 12 months is given for DAC UNIVER…
-
179, 60 53 362 D3472 D3472.201.10.20.02 07.2016 179 Sirona Dental A/S 8Operation DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL 8.3Reprocessing with DAC UNIVERSAL STANDARD båÖäáëÜ 6. Place the small lid back on the basket. 7. Close the basket. 8. Insert the BASKET lid in the cover holder of the DAC UNIVERSAL in such a way that the basket points down. …
-
116, 60 53 362 D3472 116 D3472.201.10.20.02 07.2016 5Operation DAC UNIVERSAL ADVANCED Sirona Dental A/S 5.3Reprocessing with the DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL General attachment of scaling tips 1. The scaler tip was unscrewed from the ultrasonic handpiece with the torque wrench and the tip remains in the torque wrench. 2. Ensure that the torque wre…
Sirona DAC UNIVERSAL Recommended Instructions:
CX1, CLASSIC COFFEE, UR 2 — DEVICE BRAND CODE LIST, 865511Y — Netfinity 3500 M10, MVW18MN
-
Winco
3130
Page 1 of 4 Rev E Date 07-23-12 JWC TO REDUCE THE RISK OF SEVERE INJURY OR DEATH TO PERSONS: 1. READ AND FOLLOW ALL DIRECTIONS. 2. NEVER place screen on a cart, table, stand or other objects; Screen could fall and SERIOUS INJURY MAY OCCUR. 3. NEVER place screen near open flame or over a radiator or heat register. 4. NEVER place screen in front of a door; In cas …
3130 4
-
Orliman
ONE PLUS OPL480
REF.: OPL480 | OPL481 | ONE PLUSEspañolINSTRUCCIONES DE USO Y CONSERVACIÓN Estimado cliente,Muchas gracias por su conanza en un producto Orliman. Usted adquirió un producto de calidad y de alto nivel medicinal. Por favor, lea las instrucciones atentamente. Si tienen alguna duda, pón-gase en contacto con su médico, su ortopedia especializada o con nuestro departamento de aten …
ONE PLUS OPL480 12
-
ECOPOSTURAL
C5952H
C5952HCAMILLA ELÉCTRICA / ELECTRIC TABLE / TABLE ÉLECTRIQUEINSTRUCCIONES — INSTRUCTIONS* Leer antes de utilizar por primera vez.* Read carefully before the first use.* A lire avant la première utilisation.Actualización 24/01/2017 …
C5952H 17
-
Codonics
Virtua
sáêíì~jÉÇáÅ~ä=aáëÅ=mìÄäáëÜÉêrëÉêÛë=j~åì~äsCodonics® Catalog Number VIRTUA-MNLUOctober 11, 2018Version 5.2.0Codonics, Inc.17991 Englewood DriveMiddleburg Heights, OH 44130 USA440.243.1198 Phone440.243.1334 FaxEmail [email protected] …
Virtua 198
-
Iridex
G-Probe Illuminate
CAUTION: Federal law restricts this device to sale by or on the order of a physician. IRIDEX Corporation1212 Terra Bella Ave, Mountain View, CA 94043, USAT: 650.962.8100 F: 650.962.0486 [email protected] P/N: 15007 Rev D 6/2018 1Emergo EuropePrinsessegracht 202514 AP The HagueThe NetherlandsMade in USA MYDLPIQINDICATIONSThe IRIDEX Cyclo G6™ Laser System and Probe Delivery Device …
G-Probe Illuminate 14
-
GE
ÄKTAprocess
28-9266-70 AA Sightglass cover assembly instructionGE HealthcareÄKTAprocess™ Air Trap sightglass coverAir traps with part numbers listed in the table at the end of this document must be fitted with a metal cover for added protection in case the air trap is subjected to a sudden overpressure.Follow the instructions in this document to mount the cover assembly correctly …
ÄKTAprocess 4
-
Orliman
actius ACAN08
byESACAN08#UNIACAN09#UNIANDADORESInstrucciones de uso y conservaciónORLIMAN S. L. U. C/ Ausias March, 3 — Pol. Ind. La Pobla-L´ElianaApdo. de correos 49 · C.P.: 46185La Pobla de Vallbona Valencia — España (Spain)MADE IN SPAIN EDICIÓN: 2018-07LBIACAN#8ACAN08#UNI ACAN09#UNIProfundidad640 mm 550 mmAnchura510 mm 600 mmAltura740-900 mm 790-890 mmPeso del producto4, …
actius ACAN08 4
-
OFA
dynamics Elbow Support
EllenbogenbandageElbow Supportdynamics®Die vielseitige orthopädische Versorgungorthopädische VersorgungVersatile orthopaedic careVersatile orthopaedic care545105021 | Rev. 4/03/18 Ofa AustriaWasserfeldstr. 15 A – 5020 Salzburg Tel. + 43 662 848707 Fax + 43 662 849514 [email protected] www.ofaaustria.at Ofa Bamberg GmbHLaubanger 20D – 96052 BambergTel. + 49 951 6047-0Fax + 49 951 604 …
dynamics Elbow Support 18
-
Medline
HCSM70
Pulse Oximeter General Description The measurement of oxygen saturation of arterial blood (also known as pulse oxygen saturation, usually shortened as SpO2) adopts the principles of light spectra and volume tracing. The LED emits lights with two specific wavelengths, which are selective …
HCSM70 2
Additional Information:
Popular Right Now:
Operating Impressions, Questions and Answers:
Table of Contents for Sirona DAC UNIVERSAL:
-
DAC UNIVERSAL USER MANUAL 729050 — Version 16 S 36 60 53 362 D 3472 D3472.201.01.09.02 5. Reports Sterilization To ensure sterilization, the instruments must be exposed to saturated steam at a certain temperature for a certain period of time. Sterilization method The DAC UNIVERSAL sterilises by means of saturated steam. To ensure that the saturated steam is present during the entire holding time, the pressure and the t
-
DAC UNIVERSAL USER MANUAL 729050 — Version 16 S 42 60 53 362 D 3472 D3472.201.01.09.02 7. Safety, service, warranty, certificates & approvals This chapter deals with safety, service and warranty issues. We recommend the DAC UNIVERSAL user to thoroughly read this chapter before using the DAC UNIVERSAL. 7.1 Safety It is very important that the DAC UNIVERSAL is operated as described in this manual to ensure a high safety level and high efficiency of the autoclave. In addition to this, some precautions should be considered when using the DAC UNIV
-
DAC UNIVERSAL USER MANUAL 729050 — Version 16 S 14 60 53 362 D 3472 D3472.201.01.09.02 Post processing information: 5. After the DAC UNIVERSAL has finished the cycle, allow the handpieces/turbines to cool down before handling, as they are hot after a cycle. 6. i NOTE The chuck of the instruments should after a mechanical cleaning; always additionally be lubricated according to the recommendations of the manufacturer of the i
-
DAC UNIVERSAL USER MANUAL 729050 — Version 16 S 60 53 362 D3472 3 D3472.201.01.09.02 CONTENT 1. Introduction to the DAC UNIVERSAL ………………………………………………………………………………5 1.1 Safety and technical Information ……………………………………………………………………………….. 8 1.1.1 ESD protective measures ………………………………………………………………………………………………………………
-
DAC UNIVERSAL USER MANUAL 729050 — Version 16 S 60 53 362 D3472 31 D3472.201.01.09.02 4. Run the DAC UNIVERSAL unit through a normal cleaning, lubrication and sterilization cycle. 5. Allow the chamber to cool and remove the biological indicator and/or indicator vial from the chamber. If the Spore-/PCD-Tester has been used — the biological spore test indicator and/or indicator strip is removed by unscrewing the Spore-/PCD-Tester (be careful — the Spore-/PCD-Tester is very hot after the sterilization cycle).
-
DAC UNIVERSAL USER MANUAL 729050 — Version 16 S 60 53 362 D3472 32 D3472.201.01.08.02 3.7 Using the menu system The DAC UNIVERSAL has a menu system, which gives the user different options. The different options and how to use the menu system will be explained below. To enter menu level 1, press the menu button, press Up or Down button until the desired option is selected, and press Enter to select. At each menu level, press Up/Down until the desi
-
DAC UNIVERSAL USER MANUAL 729050 — Version 16 S 60 53 362 D3472 39 D3472.201.01.09.02 or Procedure B) 1. Press the Menu/Enter button. 2. 3.
-
DAC UNIVERSAL USER MANUAL 729050 — Version 16 S 60 53 362 D3472 17 D3472.201.01.09.02 Maximum length of hand pieces and turbines: The six adapter positions on the adapter lid are each marked with dots, which symbolises the adapter-position-number on the adapter lid. On position 1 on the lid the instruments must not be longer than 145 mm/5.71 inches (total length of instrument and adapter). The maximum length on position 2-6 is 160 mm/6.30 inches (total length of instrument and
-
DAC UNIVERSAL USER MANUAL 729050 — Version 16 S 60 53 362 D3472 11 D3472.201.01.09.02 Radiated HF interfer- ence IEC 6000-4-3 3V/m 80MHz to 800MHz 1 3V/m 800MHz to 2,5GHz 1 3 Veff 3 Veff d = [1,2] P at 80 MHz to 800 MHz d = [2,3] P at 800 MHz to 2,5 GHz where P is the nominal transmitter output in watts (W) specified by the transmitter manufacturer and d is
-
DAC UNIVERSAL USER MANUAL 729050 — Version 16 S 44 60 53 362 D 3472 D3472.201.01.09.02 Appendix 2: DAC UNIVERSAL display & keyboard Display/button Description 1. Display The temperature and pressure in the autoclave chamber is constantly shown. The pressure is displayed as an absolute value. (Includes surrounding air pressure). 2. Me
-
DAC UNIVERSAL USER MANUAL 729050 — Version 16 S 60 53 362 D3472 9 D3472.201.01.09.02 0123 0123 An electrostatic discharge is: perceptible at 3,000 V or higher audible at 5,000 V or higher (cracking, crackling) visible at 10,000 V or higher (arc-over) The transient currents resulting from these discharges have a magnitude of 10 amperes. They are not hazardous for hu
Questions, Opinions and Exploitation Impressions:
You can ask a question, express your opinion or share our experience of Sirona DAC UNIVERSAL device using right now.
-
121, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 121 Sirona Dental A/S 6Cleaning and maintanance DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL båÖäáëÜ 6 Cleaning and maintanance DAC UNIVERSAL ADVANCED After every 50 cycles, the DAC UNIVERSAL will give a 2-second acoustic alarm and the display will indicate as shown on the left. This message reminds the user that …
-
170, 60 53 362 D3472 170 D3472.201.10.20.02 07.2016 8Operation DAC UNIVERSAL STANDARD Sirona Dental A/S 8.2Using the menu and display information DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL Deactivating the cleaning function For example, for reprocessing wrapped instruments with the WRAPPED lid. ✔ The DAC UNIVERSAL is switched on and ready for operation. 1. Press the «…
-
49, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 49 Sirona Dental A/S 4Installation Operating Instructions DAC UNIVERSAL 4.1Requirements of the location / Ambient conditions båÖäáëÜ 4 Installation 4.1 Requirements of the location / Ambient conditions Requirements To find a suitable location for the DAC UNIVERSAL, observe the following requirements: ● ● Open, well-ventilated area ●…
-
38, Sirona DAC UNIVERSAL 60 53 362 D3472 38 D3472.201.10.20.02 07.2016 3Product description Sirona Dental A/S 3.5Presentation of the sterilization process Operating Instructions DAC UNIVERSAL Sensitive program The following graph shows temperature (°C) and pressure (bar) during the sensitive program, with standard settings, holding time 15 min and a total cycle time of 28 min: Se…
-
230, Sirona DAC UNIVERSAL 60 53 362 D3472 230 D3472.201.10.20.02 07.2016 11Accessories Sirona Dental A/S 11.1Adapters for the STANDARD lid Operating Instructions DAC UNIVERSAL 11.1.1.11 Midwest/ISO 4/5-hole fixed connection adapter Dealing wi th adapters_Sirona_ Nitram Follow the general safety information in «General placement of instruments DAC UNIVERSAL ADVANCED [ → 99]&q…
-
125, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 125 Sirona Dental A/S 6Cleaning and maintanance DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL 6.5Drain the water tank of the DAC UNIVERSAL ADVANCED båÖäáëÜ 3. Assure the lid is attached vertically on the bottom adapter, as shown. 4. Enter the «manual water channels operations“ option by pressing the “Left”-Key fro…
-
72, Sirona DAC UNIVERSAL 60 53 362 D3472 72 D3472.201.10.20.02 07.2016 5Operation DAC UNIVERSAL ADVANCED Sirona Dental A/S 5.2Using the menu and display information DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL Indicates that the BASKET lid has been selected for the current program. The user shall confirm that this cycle is in fact being run with this lid by pressing the “Down”-Key. If not…
-
152, Sirona DAC UNIVERSAL 60 53 362 D3472 152 D3472.201.10.20.02 07.2016 7Display messages DAC UNIVERSAL ADVANCED Sirona Dental A/S 7.3Reset DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL If the alarm lamp flashes, but no acoustic alarm is given, the system has reached 3000 cycles or the last servicing was 24 months ago. Please contact your authorized dealer for serv…
-
173, 60 53 362 D3472 D3472.201.10.20.02 07.2016 173 Sirona Dental A/S 8Operation DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL 8.3Reprocessing with DAC UNIVERSAL STANDARD båÖäáëÜ General attachment Place the instruments on their respective adapters of the STANDARD lid. Depending on the instrument and the kind of adapter, you must press…
-
34, 60 53 362 D3472 34 D3472.201.10.20.02 07.2016 3Product description Sirona Dental A/S 3.3Programs and program selection Operating Instructions DAC UNIVERSAL Select DAC U. ADVANCED lid Program for STANDARD lid One of the 3 programs listed above can be selected from the menu. If nothing is selected, the standard program is used. Program for BASKET lid One of th…
-
11, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 11 Sirona Dental A/S 1General Information Operating Instructions DAC UNIVERSAL 1.6Structure of the document båÖäáëÜ 1.6 Structure of the document 1.6.1 Identification of the danger levels To prevent personal injury and material damage, please observe the warning and safety information provided in these ope…
-
243, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 243 Sirona Dental A/S 11Accessories Operating Instructions DAC UNIVERSAL 11.4Printer båÖäáëÜ –Drying – Process end – Confirmation (User confirmation) – Cycle end Not all of these phases are included for all lid types. The ones not included are printed as “Deactivated”. Printer output information without FLEX lid_Sirona_…
-
139, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 139 Sirona Dental A/S 6Cleaning and maintanance DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL 6.15Changing the O-rings on the adapters båÖäáëÜ 5. Screw the lid of the filter housing back on clockwise with the filter key. The drain filter has been replaced. 6.15 Changing the O-rings on the adapters Frequen…
-
112, Sirona DAC UNIVERSAL 60 53 362 D3472 112 D3472.201.10.20.02 07.2016 5Operation DAC UNIVERSAL ADVANCED Sirona Dental A/S 5.3Reprocessing with the DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL 5. After placing the instrument wrapped in a pouch in the wire brackets, pull the instrument lightly to make sure that it is properly attached between the brackets. 6. Insert t…
-
193, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 193 Sirona Dental A/S 9Cleaning and maintenance DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL 9.1Checking oil supply to the chamber of the DAC UNIVERSAL STANDARD båÖäáëÜ 9.1 Checking oil supply to the chamber of the DAC UNIVERSAL STANDARD Frequency: weekly or every 50 cycles (whichever comes first). I…
-
187, 60 53 362 D3472 D3472.201.10.20.02 07.2016 187 Sirona Dental A/S 8Operation DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL 8.3Reprocessing with DAC UNIVERSAL STANDARD båÖäáëÜ 8.3.5 Reprocessing with the FLEX lid DAC UNIVERSAL STANDARD The FLEX lid is used to clean and thermally disinfect ultrasonic scaler handpieces, scaler tips, airflow nozzles and sprayvit…
-
55, 60 53 362 D3472 D3472.201.10.20.02 07.2016 55 Sirona Dental A/S 4Installation Operating Instructions DAC UNIVERSAL 4.5Setting the display language on the DAC UNIVERSAL STANDARD båÖäáëÜ 3. Tighten the adapter with the 2 screws using the provided 1.5 mm Allen wrench. 4.5 Setting the display language on the DAC UNIVERSAL STANDARD We recommend that the main languag…
-
146, Sirona DAC UNIVERSAL 60 53 362 D3472 146 D3472.201.10.20.02 07.2016 7Display messages DAC UNIVERSAL ADVANCED Sirona Dental A/S 7.1Display messages and what they mean DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL 78 Valve error Please contact your authorized dealer. 79 Valve error Please contact your authorized dealer. 80 After reducing the water, the desired sterilization tempe…
-
194, 60 53 362 D3472 194 D3472.201.10.20.02 07.2016 9Cleaning and maintenance DAC UNIVERSAL STANDARD Sirona Dental A/S 9.3Checking water supply to the lids of the DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL Water is pumped through the internal hoses. 6. Interrupt this process as soon as water flows into the chamber by pressing the «Clear&qu…
-
190, 60 53 362 D3472 190 D3472.201.10.20.02 07.2016 8Operation DAC UNIVERSAL STANDARD Sirona Dental A/S 8.3Reprocessing with DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL 13. Press the «Clear»-Key. The FLEX lid will open completely. 14. Remove the FLEX lid from the holder on the DAC UNIVERSAL at the end of the cycle to allow the instruments to cool down f…
-
4, Sirona DAC UNIVERSAL 60 53 362 D3472 4 D3472.201.10.20.02 07.2016 Table of Contents Sirona Dental A/S Operating Instructions DAC UNIVERSAL 3.2.4 Back of unit ………………………………………………………………………….. 31 3.2.5 Safety ring ……………………………………………………………………………. 32 3.2.6 Water tank with si…
-
26, Sirona DAC UNIVERSAL 60 53 362 D3472 26 D3472.201.10.20.02 07.2016 3Product description Sirona Dental A/S 3.2Controls and functional elements of the DAC UNIVERSAL Operating Instructions DAC UNIVERSAL 3.2 Controls and functional elements of the DAC UNIVERSAL Display DAC UNIVERSAL ADVANCED With the DAC UNIVERSAL ADVANCED a new appearance has been introduced see „Overview DAC UNIVERSAL ADVANCED [ �…
-
186, 60 53 362 D3472 186 D3472.201.10.20.02 07.2016 8Operation DAC UNIVERSAL STANDARD Sirona Dental A/S 8.3Reprocessing with DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL The instruments can now be used again. They can also be stored for later use. After the end of the cycle, the DAC UNIVERSAL automatically reactivates cleaning and lubrication for t…
-
59, 60 53 362 D3472 D3472.201.10.20.02 07.2016 59 Sirona Dental A/S 5Operation DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL 5.1Regular activities båÖäáëÜ 5 Operation DAC UNIVERSAL ADVANCED 5.1 Regular activities Introducti on regular activiti es The DAC UNIVERSAL was installed by a service technician according to the instructions in the «Installation [ → 49]»…
-
41, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 41 Sirona Dental A/S 3Product description Operating Instructions DAC UNIVERSAL 3.5Presentation of the sterilization process båÖäáëÜ WRAPPED lid The following graph shows temperature (°C) and pressure (bar) during the standard program with the WRAPPED lid, with standard settings, holding time 3 min and a total cycle time o…
-
247, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 247 Sirona Dental A/S 13Disposal Operating Instructions DAC UNIVERSAL båÖäáëÜ 13 Disposal In accordance with Directive 2012/19/EU and national disposal regulations regarding old electrical and electronic devices, please be advised that such items must be disposed of in a special way within the European Union (EU). These regulations requi…
-
126, Sirona DAC UNIVERSAL 60 53 362 D3472 126 D3472.201.10.20.02 07.2016 6Cleaning and maintanance DAC UNIVERSAL ADVANCED Sirona Dental A/S 6.5Drain the water tank of the DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL 1. Enter the «manual water operation“ menu section. 2. Press the “Down”-Key. Water is pumped into the chamber. 3. When the chamber is 3/4 filled with water, …
-
48, 60 53 362 D3472 48 D3472.201.10.20.02 07.2016 3Product description Sirona Dental A/S 3.7Safety devices Operating Instructions DAC UNIVERSAL 3.7 Safety devices Safety valve The safety valve was developed according to the EN 4126-1 standard and calibrated according to the EN 764-7 standard. The safety valve in the DAC UNIVERSAL opens when the relative pressure in the chamber exceeds 3.5 …
-
233, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 233 Sirona Dental A/S 11Accessories Operating Instructions DAC UNIVERSAL 11.1Adapters for the STANDARD lid båÖäáëÜ 11.1.1.14 Adapter for Morita CP4 Dealing with adapters_Sirona_Nitram Follow the general safety information in «General placement of instruments DAC UNIVERSAL ADVANCED [ → 99]» or «General placement o…
-
130, 60 53 362 D3472 130 D3472.201.10.20.02 07.2016 6Cleaning and maintanance DAC UNIVERSAL ADVANCED Sirona Dental A/S 6.7Cleaning the drain filter Operating Instructions DAC UNIVERSAL 9. Wipe the sieve dry with a clean cloth. 10. Grasp the recesses of water tank with both hands again. 11. Insert the water tank back into the DAC UNIVERSAL. 12. Wriggle slightly until the water tank is firmly in p…
-
208, 60 53 362 D3472 208 D3472.201.10.20.02 07.2016 9Cleaning and maintenance DAC UNIVERSAL STANDARD Sirona Dental A/S 9.13Checking the safety ring of the DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL 9.13 Checking the safety ring of the DAC UNIVERSAL STANDARD Frequency: every 4 weeks or every 200 cycles (whichever comes first). ✔ The DAC UNIVERSAL is switched on and re…
-
89, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 89 Sirona Dental A/S 5Operation DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL 5.2Using the menu and display information DAC UNIVERSAL ADVANCED båÖäáëÜ 5.2.2.3.3.5 Oil expel Condensation and excess oil are purged from the instruments with air. Standard setting: Level 1. The maximum is level 5. Please note that oil is purged bef…
-
53, 60 53 362 D3472 D3472.201.10.20.02 07.2016 53 Sirona Dental A/S 4Installation Operating Instructions DAC UNIVERSAL 4.3Using the lubrication concentrate båÖäáëÜ 4.3 Using the lubrication concentrate The NitramOil #2 bottle is not yet mounted when the unit is delivered. The bottle must be mounted and oil pumped through the hoses before the first use of the DAC UNIVERSAL. For further …
-
162, 60 53 362 D3472 162 D3472.201.10.20.02 07.2016 8Operation DAC UNIVERSAL STANDARD Sirona Dental A/S 8.1Regular activities Operating Instructions DAC UNIVERSAL Replacement of the oil bottle ✔ The lid of the DAC UNIVERSAL water tank is open. 1. Unscrew the empty oil bottle counterclockwise and remove it from the oil dispenser. 2. Insert the new oil bottle thread down into the thread of the …
-
147, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 147 Sirona Dental A/S 7Display messages DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL 7.1Display messages and what they mean DAC UNIVERSAL ADVANCED båÖäáëÜ 84 The sterilization temperature was not maintained during sterilization. If the water level is too low, fill demineralized or distilled water (wa…
-
57, 60 53 362 D3472 D3472.201.10.20.02 07.2016 57 Sirona Dental A/S 4Installation Operating Instructions DAC UNIVERSAL 4.6Connecting to a documentation device båÖäáëÜ 4.6.1 Connecting to a documentation device DAC UNIVERSAL ADVANCED To connect a documentation device to the DAC UNIVERSAL ADVANCED, please proceed as follows: ✔ The DAC UNIVERSAL is switched on. 1. Co…
-
191, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 191 Sirona Dental A/S 8Operation DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL 8.3Reprocessing with DAC UNIVERSAL STANDARD båÖäáëÜ 21. Remove every instrument from the FLEX lid. See «Attaching instruments to the respective FLEX lid adapters [ → 238]» for detailed instructions on removing the individua…
-
209, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 209 Sirona Dental A/S 9Cleaning and maintenance DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL 9.14Changing the drain filter båÖäáëÜ 9.14 Changing the drain filter Frequency: every 4 weeks or every 200 cycles (whichever comes first). ✔ The DAC UNIVERSAL is switched off and cooled down. ✔ You are wear…
-
95, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 95 Sirona Dental A/S 5Operation DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL 5.2Using the menu and display information DAC UNIVERSAL ADVANCED båÖäáëÜ 5.2.4 Symbol definitions DAC UNIVERSAL ADVANCED Refer to this list to see the definition of symbols used on the display of the DAC UNIVERSAL. For…
-
17, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 17 Sirona Dental A/S 2Safety instructions Operating Instructions DAC UNIVERSAL 2.2Caution symbols on the device båÖäáëÜ 2.2 Caution symbols on the device Meaning: Warning of danger This symbol is found on the back of the DAC UNIVERSAL next to: ● The COM PORT: External computer hardware that is connected to this…
-
225, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 225 Sirona Dental A/S 11Accessories Operating Instructions DAC UNIVERSAL 11.1Adapters for the STANDARD lid båÖäáëÜ 11.1.1.6 Adapter for Sirona T1 CLASSIC/TE Dealing with adapters_Sirona_Nitram Follow the general safety information in «General placement of instruments DAC UNIVERSAL ADVANCED [ → 99]&…
-
144, 60 53 362 D3472 144 D3472.201.10.20.02 07.2016 7Display messages DAC UNIVERSAL ADVANCED Sirona Dental A/S 7.1Display messages and what they mean DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL 70 The pressure in the chamber does not drop quickly enough after the pressure test. Drain water from the chamber. Clean or change the drain filt…
-
145, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 145 Sirona Dental A/S 7Display messages DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL 7.1Display messages and what they mean DAC UNIVERSAL ADVANCED båÖäáëÜ 74 When heating for sterilization, a temperature of 120°C / 248°F was not reached within 2.5 minutes. Increase the level of the «Ambien…
-
182, 60 53 362 D3472 182 D3472.201.10.20.02 07.2016 8Operation DAC UNIVERSAL STANDARD Sirona Dental A/S 8.3Reprocessing with DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL 8.3.4 Reprocessing with the WRAPPED lid DAC UNIVERSAL STANDARD The WRAPPED lid is used for reprocessing wrapped rotating handpieces and wrapped solid instruments. WRAPPED lid …
-
129, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 129 Sirona Dental A/S 6Cleaning and maintanance DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL 6.6Cleaning the water tank båÖäáëÜ 5. Clean the water tank with warm water (maximum 50ºC / 122ºF). 6. Clean the filter with warm water (maximum 50ºC / 122ºF). 7. Clean the float in the water t…
-
80, Sirona DAC UNIVERSAL 60 53 362 D3472 80 D3472.201.10.20.02 07.2016 5Operation DAC UNIVERSAL ADVANCED Sirona Dental A/S 5.2Using the menu and display information DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL Rinse cycles 5 Rinse cycles 6 Oil expel Oil expel 1 Oil expel 2 Oil expel 3 Oil expel 4 Oil expel 5 Expel: Expel 0 Expel 1 Expel 2 Expel 3 Set date Choose date Choose month Level 1 Leve…
-
238, Sirona DAC UNIVERSAL 60 53 362 D3472 238 D3472.201.10.20.02 07.2016 11Accessories Sirona Dental A/S 11.2Adapters for the FLEX lid Operating Instructions DAC UNIVERSAL 11.2 Adapters for the FLEX lid FLEX lid adapters_Sirona 11.2.1 Attaching instruments to the respective FLEX lid adapters Tip: If there are many instruments to be reprocessed or to ensure the instruments will be avai…
-
13, 60 53 362 D3472 D3472.201.10.20.02 07.2016 13 Sirona Dental A/S 1General Information Operating Instructions DAC UNIVERSAL 1.8Intended use båÖäáëÜ 1.8 Intended use The DAC UNIVERSAL is intended to clean, lubricate and steam sterilize dental handpieces. Dental handpieces is defined as rotating instruments split up into the categories straight handpieces, contra angles …
-
120, Sirona DAC UNIVERSAL 60 53 362 D3472 120 D3472.201.10.20.02 07.2016 5Operation DAC UNIVERSAL ADVANCED Sirona Dental A/S 5.3Reprocessing with the DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL 22. Remove every instrument from the FLEX lid. See «Attaching instruments to the respective STANDARD lid adapters [ → 220]» for detailed instructions on removing the individual in…
-
76, 60 53 362 D3472 76 D3472.201.10.20.02 07.2016 5Operation DAC UNIVERSAL ADVANCED Sirona Dental A/S 5.2Using the menu and display information DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL The error code is displayed in the top left hand corner; Low water. Make sure the DAC UNIVERSAL is supplied with water from the water treatment system. For further troubleshoo…
-
115, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 115 Sirona Dental A/S 5Operation DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL 5.3Reprocessing with the DAC UNIVERSAL ADVANCED båÖäáëÜ 5.3.5 Reprocessing with the FLEX lid DAC UNIVERSAL ADVANCED The FLEX lid is used to clean and thermally disinfect ultrasonic scaler handpieces, scaler tips, airflow nozzles and sprayvit noz…
-
82, 60 53 362 D3472 82 D3472.201.10.20.02 07.2016 5Operation DAC UNIVERSAL ADVANCED Sirona Dental A/S 5.2Using the menu and display information DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL 70% 80% 90% 100% Automatic water Automatic water ON Automatic water OFF History Cycles Cycle history Service Next service Errorbuffer Errors 1-3 Errors 4-6 Errors 7-9 Level 1 Leve…
-
30, 60 53 362 D3472 30 D3472.201.10.20.02 07.2016 3Product description Sirona Dental A/S 3.2Controls and functional elements of the DAC UNIVERSAL Operating Instructions DAC UNIVERSAL 3.2.3.2 Display and keyboard DAC UNIVERSAL STANDARD Display/button Description A: Display The temperature and pressure in the chamber are displayed constantly. The pressure is indicated as an absolute value (…
-
81, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 81 Sirona Dental A/S 5Operation DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL 5.2Using the menu and display information DAC UNIVERSAL ADVANCED båÖäáëÜ Choose year Confirm date setting Set time 12 hour clock Choose AM/PM hours Choose minutes Confirm time setting 24 hour clock Choose hours Choose minutes…
-
71, 60 53 362 D3472 D3472.201.10.20.02 07.2016 71 Sirona Dental A/S 5Operation DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL 5.2Using the menu and display information DAC UNIVERSAL ADVANCED båÖäáëÜ 5.2.1 Display information DAC UNIVERSAL ADVANCED The display will show the current system status. Below examples of these status messages are shown. P…
-
32, 60 53 362 D3472 32 D3472.201.10.20.02 07.2016 3Product description Sirona Dental A/S 3.2Controls and functional elements of the DAC UNIVERSAL Operating Instructions DAC UNIVERSAL 3.2.5 Safety ring The DAC UNIVERSAL is equipped with a safety ring that serves as a cycle confirmation tool and prevents users from pinching their fingers when closing the l…
-
44, Sirona DAC UNIVERSAL 60 53 362 D3472 44 D3472.201.10.20.02 07.2016 3Product description Sirona Dental A/S 3.6Electromagnetic compatibility Operating Instructions DAC UNIVERSAL 3.6 Electromagnetic compatibility In order to prevent possible risks caused by electromagnetic interference, no medical electrical or other equipment should be used near the DAC UNIVERSAL. The DAC UNIVERSAL com…
-
178, Sirona DAC UNIVERSAL 60 53 362 D3472 178 D3472.201.10.20.02 07.2016 8Operation DAC UNIVERSAL STANDARD Sirona Dental A/S 8.3Reprocessing with DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL 3. Open the snap lock on the basket and take the small lid off the basket. 4. Put maximum 5 solid instruments into the basket, if possible with the sharp tip down. The solid instruments must be placed…
-
169, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 169 Sirona Dental A/S 8Operation DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL 8.2Using the menu and display information DAC UNIVERSAL STANDARD båÖäáëÜ 8.2.3 Standard settings and Resetting the menu system DAC UNIVERSAL STANDARD Standard settings Only for Japan: ambient temperature sensitivity is set at 4. Ple…
-
75, 60 53 362 D3472 D3472.201.10.20.02 07.2016 75 Sirona Dental A/S 5Operation DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL 5.2Using the menu and display information DAC UNIVERSAL ADVANCED båÖäáëÜ The error code is displayed in the top left hand corner; Displays when the water quality is insufficient. The water quality needs to be 3.0 μS/cm or better. For further…
-
245, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 245 Sirona Dental A/S 12Consumables, spare parts, and tools Operating Instructions DAC UNIVERSAL båÖäáëÜ 12 Consumables, spare parts, and tools List of spare parts for maintenance of adapters REF Adapters for highspeed handpieces ME Spare O- ring REF Spare O-ring designation 6051713 Adapter for BienAir UNIFIX 1 6079839 O…
-
35, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 35 Sirona Dental A/S 3Product description Operating Instructions DAC UNIVERSAL 3.4Technical data båÖäáëÜ 3.4 Technical data Manufacturer: Sirona Dental A/S Model: DAC UNIVERSAL Dimensions (H/W/D): Opened: 53 cm / 38 cm / 37 cm Closed with STANDARD lid: 35 cm / 38 cm / 37 cm Materials: Anodized aluminum, plastic Weight: (net witho…
-
33, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 33 Sirona Dental A/S 3Product description Operating Instructions DAC UNIVERSAL 3.3Programs and program selection båÖäáëÜ 3.2.7 NitramOil #2 lubrication concentrate The NitramOil #2 lubrication concentrate is for lubricating the transmission channels of the instruments and is screwed into the oil dispenser in the DAC …
-
24, 60 53 362 D3472 24 D3472.201.10.20.02 07.2016 3Product description Sirona Dental A/S 3.1Functions Operating Instructions DAC UNIVERSAL Sterilization The wrapped instruments are sterilized for 3 minutes at 134°C (273°F). Functions WRAPPED lid with ADVANCED For instructions for the sterilization cycle for wrapped instruments see «Reprocessing with …
-
227, 60 53 362 D3472 D3472.201.10.20.02 07.2016 227 Sirona Dental A/S 11Accessories Operating Instructions DAC UNIVERSAL 11.1Adapters for the STANDARD lid båÖäáëÜ 11.1.1.8 Adapter for Castellini CERAMIC FREEDOM Dealing with adapters_Sirona_Nitram Follow the general safety information in «General placement of instruments DAC UNIVERSAL ADVANCED [ → 99]» or…
-
117, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 117 Sirona Dental A/S 5Operation DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL 5.3Reprocessing with the DAC UNIVERSAL ADVANCED båÖäáëÜ 5.3.5.2 Start cycle on the FLEX lid DAC UNIVERSAL ADVANCED ✔ All settings of the DAC UNIVERSAL were preset according to «Regular activities [ → 59]». …
-
19, 60 53 362 D3472 D3472.201.10.20.02 07.2016 19 Sirona Dental A/S 2Safety instructions Operating Instructions DAC UNIVERSAL 2.5Instruction of practice personnel båÖäáëÜ 2.5 Instruction of practice personnel The personnel responsible for operation and maintenance of the DAC UNIVERSAL must be instructed in the safe operation of the DAC UNIVERSAL by a service …
-
143, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 143 Sirona Dental A/S 7Display messages DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL 7.1Display messages and what they mean DAC UNIVERSAL ADVANCED båÖäáëÜ 64 The safety system was activated. Please contact your authorized dealer. 65 The safety system is defective. Please contact your authorized dealer. 66 The temperature…
-
248, tÉ=êÉëÉêîÉ=íÜÉ=êáÖÜí=íç=ã~âÉ=~åó=~äíÉê~íáçåë=ïÜáÅÜ=ã~ó=ÄÉ=êÉèìáêÉÇ=ÇìÉ=íç=íÉÅÜåáÅ~ä=áãéêçîÉãÉåíëK «=páêçå~=aÉåí~ä=^Lp=OMNS péê~ÅÜÉW ÉåÖäáëÅÜ mêáåíÉÇ=áå=aÉåã~êâ aPQTOKOMNKNMKOMKMO MTKOMNS ûKJkêKW= ahMRNT páêçå~=aÉåí~ä=^Lp oÜç=NM ahJUPUO=eáåå…
-
98, Sirona DAC UNIVERSAL 60 53 362 D3472 98 D3472.201.10.20.02 07.2016 5Operation DAC UNIVERSAL ADVANCED Sirona Dental A/S 5.3Reprocessing with the DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL 5.3 Reprocessing with the DAC UNIVERSAL ADVANCED 5.3.1 Switching ON/OFF/Standby mode DAC UNIVERSAL ADVANCED Switch-on The DAC UNIVERSAL will power up and heat the steam generator once the DAC UNIVERSAL …
-
229, 60 53 362 D3472 D3472.201.10.20.02 07.2016 229 Sirona Dental A/S 11Accessories Operating Instructions DAC UNIVERSAL 11.1Adapters for the STANDARD lid båÖäáëÜ 11.1.1.10 Adapter for NSK QDJ Dealing with adapters_Sirona_Nitram Follow the general safety information in «General placement of instruments DAC UNIVERSAL ADVANCED [ → 99]» or «General placement …
-
200, Sirona DAC UNIVERSAL 60 53 362 D3472 200 D3472.201.10.20.02 07.2016 9Cleaning and maintenance DAC UNIVERSAL STANDARD Sirona Dental A/S 9.6Cleaning the water tank Operating Instructions DAC UNIVERSAL 7. Clean the float in the water tank carefully with a moist cloth. 8. Wipe the tank dry with a clean cloth. 9. Wipe the sieve dry with a clean cloth. 10. Grasp the recesses of water …
-
103, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 103 Sirona Dental A/S 5Operation DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL 5.3Reprocessing with the DAC UNIVERSAL ADVANCED båÖäáëÜ 14. Remove the STANDARD lid from the holder on the DAC UNIVERSAL at the end of the cycle to allow the instruments to cool down faster. 15. Place the STANDARD lid on the counter or put it …
-
212, Sirona DAC UNIVERSAL 60 53 362 D3472 212 D3472.201.10.20.02 07.2016 10Display messages DAC UNIVERSAL STANDARD Sirona Dental A/S 10.1Display messages and what they mean DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL 10.1 Display messages and what they mean DAC UNIVERSAL STANDARD The display messages have the following meanings: Code no. Description Corrective ac…
-
172, Sirona DAC UNIVERSAL 60 53 362 D3472 172 D3472.201.10.20.02 07.2016 8Operation DAC UNIVERSAL STANDARD Sirona Dental A/S 8.3Reprocessing with DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL 8.3.2 Reprocessing with the STANDARD lid DAC UNIVERSAL STANDARD The STANDARD lid is used for reprocessing unwrapped rotating handpieces. Tip: If there are many instruments to be reprocessed or to ensure the i…
-
174, Sirona DAC UNIVERSAL 60 53 362 D3472 174 D3472.201.10.20.02 07.2016 8Operation DAC UNIVERSAL STANDARD Sirona Dental A/S 8.3Reprocessing with DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL 8.3.2.2 Start cycle on the STANDARD lid DAC UNIVERSAL STANDARD ✔ All settings of the DAC UNIVERSAL were preset according to «Regular activities [ → 153]». ✔ The DAC UNIVERSAL is switched on …
-
156, 60 53 362 D3472 156 D3472.201.10.20.02 07.2016 8Operation DAC UNIVERSAL STANDARD Sirona Dental A/S 8.1Regular activities Operating Instructions DAC UNIVERSAL 2. Attach the STANDARD indicator holder in the groove on the center post of the STANDARD lid. 3. Start a cycle. See «Reprocessing with the STANDARD lid DAC UNIVERSAL STANDARD [ → 172]» for the detailed procedure.…
-
63, Sirona DAC UNIVERSAL 60 53 362 D3472 D3472.201.10.20.02 07.2016 63 Sirona Dental A/S 5Operation DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL 5.1Regular activities båÖäáëÜ 2. Attach the STANDARD indicator holder to one of the three wire brackets of the WRAPPED lid so the brackets of the indicator holder point to the center of the lid. 3. Press the STANDARD indicator holder towa…
-
36, Sirona DAC UNIVERSAL 60 53 362 D3472 36 D3472.201.10.20.02 07.2016 3Product description Sirona Dental A/S 3.4Technical data Operating Instructions DAC UNIVERSAL The DAC UNIVERSAL complies with the requirements of the RoHS Directive 2011/65/EU. Symbols on the DAC UNIVERSAL packaging/external carton This way up Keep away from rain Handle with care Use no hooks Temperature li…
-
86, Sirona DAC UNIVERSAL 60 53 362 D3472 86 D3472.201.10.20.02 07.2016 5Operation DAC UNIVERSAL ADVANCED Sirona Dental A/S 5.2Using the menu and display information DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL Drain chamber Drains the chamber through the drain outlet. Select this function, if there’s residue in the chamber that needs to be drained. An empty lid must b…
-
96, 60 53 362 D3472 96 D3472.201.10.20.02 07.2016 5Operation DAC UNIVERSAL ADVANCED Sirona Dental A/S 5.2Using the menu and display information DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL Cycle. Progress bar. STANDARD lid STANDARD lid is selected. BASKET lid BASKET lid is selected. FLEX lid FLEX lid is selected. WRAPPED lid WRAPPED lid is selected. Press sa…
-
221, 60 53 362 D3472 D3472.201.10.20.02 07.2016 221 Sirona Dental A/S 11Accessories Operating Instructions DAC UNIVERSAL 11.1Adapters for the STANDARD lid båÖäáëÜ 11.1.1.2 Adapter for KaVo MULTIflex Dealing with adapters_Sirona_Nitram Follow the general safety information in «General placement of instruments DAC UNIVERSAL ADVANCED [ → 99]» or «General place…
-
218, 60 53 362 D3472 218 D3472.201.10.20.02 07.2016 11Accessories Sirona Dental A/S Operating Instructions DAC UNIVERSAL 11 Accessories Accessories_Sirona with FLEX Installation kit, syphon_Sirona Installation kit, Syphon (REF 64 33 267) includes: ● Syphon with direct connection ● Manometer ● Water filter ● Holder for air filter ● Filter for filter housing (pack of 6)…